PRODUCTS >> DRILLING FLUIDS & OIL WELL >> Electrical Stability / Resistivity Tester

Electrical Stability Tester EST-23E

Date:2020-05-10 13:52:01 Clicks:536
Electrical Stability Tester EST-23E


Electrical Stability Tester (EST)

The electrical stability (ES) of an oil-based drilling fluid is related to the stability of the emulsion and oil-wetting capability.

The Electrical Stability Tester for use with oil-based drilling fluids.  Mud engineers depend on the electrical stability test to monitor a fluid’s emulsion stability and oil-wetting capability. The new tester is simpler to operate and can be used in the lab or in the field anywhere in the world.

The Electrical Stability Tester (EST) conforms to the test procedure described in  API Recommended Practice 13B-2.  It operates on external power or two nine volt batteries.  The EST is calibrated in peak volts which is the maximum voltage that the fluid experiences between the two electrodes. Peak voltage may be converted to Root Mean Square (RMS) voltage by multiplying the peak voltage by 0.7071.

Features:

Internal power supply and built in AC power plug allow universal operation

Power requirements - 100 to 240 VAC, 50/60 Hz, 10 watts

Self-calibration for immediate performance checks

Large backlit display is easy to read, even in low light

 

Components:

1.      Probe with Cable

2.   Calibration Standard for ES Meter.

3.      Switch - Push Button - On / Off

4.      Switch Boot

 

Electrical Stability Tester (EST) Specifications

Part No.

Model

Description & Technical Parameters

Remark

147-50

EST-23E

Output Frequency:340 ± 2 Hz (0-2000v)

Output wave form :Sinusoidal

Output voltage range:0 to 2025 V ± 25 V peak to ground (1432 RMS)

Breakdown peak output current :61 micro-amps (1.59mm)

Voltage ramp rate :150 ± 10 V/second

Power supply :AC 230V, Internal 27 V batteries

Battery life :Approximately 500 tests

Dimensions:25*15*12cm (9.8in*5.9*4.7in)

Complete Weight 11 lbs (5 kg)

Accuracy:Cal. Standard; ± 10 V - sample, ± 3% of reading

Repeatability:Cal. Standard; ± 5 V - sample, ± 2% of reading

Operating temperature range. :32° to 122°F (0° to 50°C)